* Suitable buffer ions for UV studies in the range of 260 to 190 nm include phosphate, Tris, acetate and borate. These buffers are suitable for pH values ranging from 5 to 10 and should be chosen in accordance to your needs.
* 10 mM of buffer is generally a good starting point if you plan to work with cuvettes of 1 mm long. For longer path cuvettes the concentration of the buffer should be reduced in accordance.
* The sample should be solubilized and as pure as possible to avoid experimental artifacts.
* The sample concentration is determined by the pathlength of the cuvette. 0.1 mg/ml of peptide or protein is a good starting point to work with 1 mm long cuvette.
* Sample volume is function of the geometry of the cuvette : 200 microL is the minimum volume needed to work with the 1 mm long cuvettes.
* Working with thinner cuvettes and highly concentrated sample is possible and can help to reduce contribution of the buffer in the spectra.
* ...